65
ZX-E
Operation Manual
Section 4 Detecting Bottom Deadpoint
Section 4
MAIN APPLICATIONS AND SETTING METHODS
Setting Measurement Timing
The hold function is used to detect the bottom deadpoint.
To ignore bounding when measuring during press operation, specify a time delay from
the timing signal to when sampling starts.
When the timing signal cannot be input from the device, set a self-down trigger.
Refer to Section 5 Detailed Settings for details on settings.
Using Hold Functions, p. 81
Measured value
Self-trigger level
The bottom value is held.
Timing input
Delay
time
Sampling
period
Z166-E1-02A.book Page 65 Friday, June 3, 2005 4:18 PM