Section 4 Measuring Thickness
Section 4
MAIN APPLICATIONS AND SETTING METHODS
Setting Expressions
Position the reference sample and set the expression to calculate the thickness of the
reference sample.
The settings are made on the CH2 Amplifier Unit. Select THICK as the expression type
and enter the thickness (T) of the reference sample.
When the thickness is entered, the positional relationship between the Sensor Heads
at that point will be registered. The thickness is measured based on the positional rela-
tionship of the Sensor Heads.
Refer to Section 6 Auxiliary Functions for details on settings.
Performing Calculations, p. 116
Setting Tolerance Judgement Values
Set the upper and lower limits (HIGH and LOW thresholds) for the thickness for a PASS
(OK) judgement.
The HIGH, PASS, and LOW judgement results will be output based on the threshold
values set here.
Refer to Section 5 Detailed Settings for details operations.
Inputting Threshold Values Directly, p. 98
Measurement result Judgement
Measurement result > HIGH threshold HIGH
LOW threshold
≤ measurement result ≤ HIGH threshold PASS
LOW threshold > Measurement result LOW