8-4 Operator’s Manual – OPTI CCA-TS2
8 DIAGNOSTICS AND TROUBLESHOOTING
<WARNING! - Check Pump>
The peristaltic pump is getting worn.
• Remove the cassette.
• Retry with a new cassette.
• Change the peristaltic pump cartridge
(See Section 7.4.1).
<WARNING! - SRC expires soon!>
The SRC will expire in four weeks.
• Press
to continue. Make sure you
have more SRCs on hand or ordered.
<WARNING! - New SRC Lot>
This message is displayed to alert the user that a new
SRC lot is being used.
• Press
to continue measurement with
the new SRC.
<WARNING! - Hb Calibration expires soon!>
The Hb calibration will expire in four weeks.
• Press
to continue.
<WARNING! - Hb Calibrator expires soon!>
The Hb Calibrator cassette will expire in four weeks.
• Press
to continue. Make sure you
have another Calibrator Cassette on hand or
ordered.
<WARNING! - New tHb Calibrator>
This message is displayed to alert the user that a new
tHb calibrator is being used.
• Press
to continue measurement with
the new tHb calibrator.