2-4 Operator’s Manual – OPTI CCA-TS2
2 SETUP
When prompted (Fig. 2-7), insert the gas
bottle in its housing and turn clockwise until
ngertight (Fig. 2-8). Press
.
NOTE: Thegasbottleexpires6monthsafter
installationorafterexceedingthe
labeledexpirationdate,whichever
comesrst.
NOTE: Thebarcodecontainsexpiration
information.Fourweekspriorto
expirationofthegasbottle,theOPTI
CCA-TS2willalerttheoperatoronce,
asaremindertoorderareplacement
gasbottle.
• When this display appears (Fig. 2-9), press
to install a new gas bottle.
NOTE: Ifaftertheinitialinstallationyouneed
toremoveagasbottleandreinstall
thesamebottle,respond 
tothe<New Gas Bottle?>prompt.
Thenextscreenwillpromptyouto
enterthenumberofweeksinservice
usingthenumerickeypad(Seesection
7.5.3).Hereyoumayreferbacktothe
installationdate,whichwasrecordedon
thegasbottle.
The OPTI CCA-TS2 will now begin to warm up
and perform a gas purge, which will be indicated
by a progress bar displayed on the screen.
Once the warm-up is complete, the <Ready>
display appears (Fig. 2-10).
Fig. 2-8 Insert Gas Bottle
Fig. 2-10 Ready screen
Fig. 2-7 Insert Gas Bottle
Fig. 2-9 New Gas Bottle