EasyManua.ls Logo

OPTi CCA-TS2 - Page 31

OPTi CCA-TS2
299 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2-6 Operator’s Manual – OPTI CCA-TS2
2 SETUP
6. Performing tHb Calibration
The tHb Calibrator Cassette should be run prior
to patient testing when rst setting up your
analyzer. The tHb calibrator should then be run
every three months. Your OPTI CCA-TS2 will
remind you when the tHb calibration is due.
The tHb Calibrator Cassette can be found in
the storage compartment in the back of your
analyzer.
• In the <Ready> display, scan the bottom
bar code on the calibrator cassette package
by holding it 2-3 inches (5-8 cm) from the
bar code scanner located on the bottom
right-hand corner of the analyzer (Fig. 2-14).
NOTE: AtHbcalibrationcanalsoberun
fromtheQCmenubypressing<QC
Manager>QC>tHb Calibrator>instead
ofscanningthebarcodeinthe<Ready>
screen.
• The red line from the bar code scanner
should cover the entire bar code.
• A beep and a green status light indicates a
valid bar code.
• If <Non Secure User ID Entry> is
enabled in the security settings (see Section
3.2.3), you will be asked to enter the user ID
(Fig. 2-15). Depending on security settings,
user access to running Hb calibrators may be
restricted.
NOTE: Bar-codeduserIDsmaybeenteredfromthis
screenusingthebarcodescanner.
• A warning will be displayed the rst time a
new tHb Calibrator lot is used (Fig. 2-16).
Press <Continue>.
Fig. 2-15 Enter User ID
Fig. 2-14 Scan Bar Code
Fig. 2-16 New Calibrator

Table of Contents