Chapter 4 - Technical Specification
4.3.1 Guidance and manufacturer declaration - electromagnetic
emissions
TABLE 10: List of electromagnetic emission compliance
Emissions test
IEC60601
compliance
test level
Electromagnetic environment-guidance
RF emissions
CISPR 11
Group 1
The P200T uses RF energy only for its internal function. Therefore,
its RF emissions are very low and are NOT likely to cause any
interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
The P200T is suitable for use in all establishments, including
domestic establishments and those directly connected to the public
low-voltage power supply network that supplies buildings used for
domestic purposes.
Harmonic emissions
IEC61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC61000-3-3
Pst
Dmax
4.3.2 Guidance and manufacturer declaration – electromagnetic
immunity
TABLE 11: List of electromagnetic immunity compliance
Immunity test
IEC60601 compliance test
level
Electromagnetic environment-guidance
Electrostatic
discharge
IEC61000-4-2
±6 kV contact
±8 kV air
Floors should be wood, concrete or ceramic tile. If floors
are covered with synthetic material, the relative humidity
should be at least 30%.
Electrical fast
transient/burst
IEC61000-4-4
±2 kV for power supply lines
±1 kV for input / output lines
Mains power quality should be that of a typical
commercial or hospital environment.
Surge
IEC61000-4-5
±1 kV line to line
±2 kV line to earth
Mains power quality should be that of a typical
commercial or hospital environment.
Page 56 of 65 Part Number: G102748/9GSE