Error message Error no. Contents Cause Possible action
Memory Error (None) Unspecified memory
error in control mod
ules
e.g.defective plug
connection, cable
break
Defective control
module
Contact qualified
personnel
PM Memory Error (None) Memory error in con
troller
e.g.defective plug
connection, cable
break
Defective controller
Check all plug con
nections and cables
on controller
If error persists, con
tact qualified person
nel
Bad Cable (None) Error on the commu
nication cables
between control
modules
e.g.defective plug
connection, cable
break
Check all plug con
nections and cables
on the control mod
ules (incl. controller)
If error persists, con
tact qualified person
nel
Module Error (None) Error in control mod
ules
e.g.unspecified error
in displayed control
module
Displayed control
module defective
Check all plug con
nections and cables
If error persists, con
tact qualified person
nel
System Error (None) System error Non-attributable error
in control modules
(primarily from third-
party providers)
Check all plug con
nections and cables;
disconnect/test any
modules from third-
party providers in
succession if applic
able
If error persists, con
tact qualified person
nel
SID Detached (None) Special control dis
connected
Special control dis
connected from the
connection module of
the TEN°LCD mod
ule (sip and puff con
trol)
Defective special
control connection on
the connection mod
ule of the TEN°LCD
module
Check all plug con
nections and cables
between special con
trol connection and
sip and puff control;
check programming
parameters
If error persists, con
tact qualified person
nel
1E07 External on/off switch
disconnected from
control panel
Switch Detached
1E08 External profile/mode
switch disconnected
from control panel
e.g.defective cable
connection to control
panel
Defective switch or
control panel
Check cable connec
tion and switching
function of the extern
al switch
If error persists, con
tact qualified person
nel
Gone to Sleep (None) Control device
switches to sleep
mode
Extended inactivity
while control device
is switched on
Switch control device
off/on
Charging (None) Battery charger is
connected
Battery charger is
connected to char
ging receptacle
Disconnect battery
charger prior to driv
ing
114
Maintenance and repair
490E75=1_C

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Otto Bock 490E75 1 C Series and is the answer not in the manual?

Otto Bock 490E75 1 C Series Specifications

General IconGeneral
BrandOtto Bock
Model490E75 1 C Series
CategoryWheelchair
LanguageEnglish

Related product manuals