EasyManuals Logo

Otto Bock 490E75 1 C Series User Manual

Otto Bock 490E75 1 C Series
128 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #115 background imageLoading...
Page #115 background image
Error message Error no. Contents Cause Possible action
JS Static Timeout (None) Joystick holding time
exceeded
Excessive joystick
deflection detected
(control device stops
drive system to pre
vent potential dam
age to the motors)
Defective joystick
Switch control device
off/on
If error persists:
replace control pan
el/joystick; contact
qualified personnel
1E0D Short circuit on
external on/off switch
Switch Short
1E0C Short circuit on
external profile/mode
switch
e.g.defective cable
connection to control
panel
Defective switch or
control panel
Check cable connec
tion and switching
function of the extern
al switch
If error persists, con
tact qualified person
nel
Additional error messages for model equipped with a gyro module
Error message Error no. Contents Cause Possible action
Gyro Disconnected (None) Electronic track sta
biliser (gyro module)
disconnected from
controller
e.g.defective cable
connection to con
troller
Defective gyro mod
ule
Contact qualified
personnel
Gyro Prf Active (None) Electronic track sta
biliser (gyro module)
is activated
Electronic track sta
biliser (gyro module)
is disrupted by the
movement of a differ
ent means of trans
portation (e.g.train,
ship) in standby
Defective gyro mod
ule
When driving/stand
ing in means of trans
portation, select the
“No Assist” driving
profile for safety reas
ons
If error persists, con
tact qualified person
nel
8.3.4 Attendant control error overview
VR2 attendant control
Flashing LED Error/warning Cause Possible measure
Control unit fault Fault in the control unit Read error on the control panel and
take corresponding action (see
"Wheelchair control unit error over
view" table)
Attendant control fault Defective attendant control Switch off attendant control and turn
control panel off/on
Joystick error Attendant control joystick
not in home position when
the unit is turned on
Switch off attendant control and move
joystick to home position before
switching the unit on
Switch control panel off/on
Communication error
between the attendant con
trol (joystick) and the con
trol panel/controller
Defective cable, loose plug
connection
Check cabling / plug connections
R-Net attendant control
Flashing LED Error/warning Cause Possible measure
Control unit fault Fault in the control unit Read error on the control panel and
take corresponding action (see
"Wheelchair control unit error over
view" table)
115490E75=1_C
Maintenance and repair

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Otto Bock 490E75 1 C Series and is the answer not in the manual?

Otto Bock 490E75 1 C Series Specifications

General IconGeneral
BrandOtto Bock
Model490E75 1 C Series
CategoryWheelchair
LanguageEnglish

Related product manuals