Error message Error no. Contents Cause Possible action
JS Static Timeout (None) Joystick holding time
exceeded
Excessive joystick
deflection detected
(control device stops
drive system to pre
vent potential dam
age to the motors)
Defective joystick
Switch control device
off/on
If error persists:
replace control pan
el/joystick; contact
qualified personnel
1E0D Short circuit on
external on/off switch
Switch Short
1E0C Short circuit on
external profile/mode
switch
e.g.defective cable
connection to control
panel
Defective switch or
control panel
Check cable connec
tion and switching
function of the extern
al switch
If error persists, con
tact qualified person
nel
Additional error messages for model equipped with a gyro module
Error message Error no. Contents Cause Possible action
Gyro Disconnected (None) Electronic track sta
biliser (gyro module)
disconnected from
controller
e.g.defective cable
connection to con
troller
Defective gyro mod
ule
Contact qualified
personnel
Gyro Prf Active (None) Electronic track sta
biliser (gyro module)
is activated
Electronic track sta
biliser (gyro module)
is disrupted by the
movement of a differ
ent means of trans
portation (e.g.train,
ship) in standby
Defective gyro mod
ule
When driving/stand
ing in means of trans
portation, select the
“No Assist” driving
profile for safety reas
ons
If error persists, con
tact qualified person
nel
8.3.4 Attendant control error overview
VR2 attendant control
Flashing LED Error/warning Cause Possible measure
Control unit fault Fault in the control unit Read error on the control panel and
take corresponding action (see
"Wheelchair control unit error over
view" table)
Attendant control fault Defective attendant control Switch off attendant control and turn
control panel off/on
Joystick error Attendant control joystick
not in home position when
the unit is turned on
Switch off attendant control and move
joystick to home position before
switching the unit on
Switch control panel off/on
Communication error
between the attendant con
trol (joystick) and the con
trol panel/controller
Defective cable, loose plug
connection
Check cabling / plug connections
R-Net attendant control
Flashing LED Error/warning Cause Possible measure
Control unit fault Fault in the control unit Read error on the control panel and
take corresponding action (see
"Wheelchair control unit error over
view" table)
115490E75=1_C
Maintenance and repair