6.5.3 Point Measurement
The Point measurement measuring method starts recording of the
measured values through the triggering of a defined trigger. The record-
ing includes the respective, current measured value of all the active
channels at the time of triggering.
The table below contains information on the triggers available:
Designation Description
Keystroke Recording of a measuring point by tapping
on the corresponding button
D-IN falling slope Recording of a measuring point on transition
of a digital signal from High to Low
D-IN rising slope Recording of a measuring point on transition
of a digital signal from Low to High
Channel alarm Recording of a measuring point if defined
alarm values occur
Channel warning Recording of a measuring point if defined
warning values occur
After selecting the Point measurement measuring method, the device
displays a predefined trigger to start the measurement.
1 Tap on the
button.
Ä The configuration window to select the trigger opens.
2 Select a trigger source according to your application.
208
The Parker Service Master CONNECT V1.0/04/20
Operation