6.5.4 Trigger
The Trigger measuring method starts recording of the measured val-
ues through the triggering of a defined trigger.
Recording is stopped automatically after the defined storage time has
expired.
After selecting the Trigger measuring method, the device displays a
predefined trigger and corresponding parameters to complete the mea-
surement.
1 Tap on the
button.
Ä The configuration window opens in which to define the param-
eters.
2 Define the parameters and select a trigger according to your ap-
plication.
The following parameters are available:
Designation Description
Storage interval Set the storage interval of the measurement in
the format h : m : s : ms
Storage time Set the storage time for the measurement in
the format d : h : m : s When the storage time
has expired, the measurement stops automati-
cally
Recording MIN/MAX If this option is activated, the Min and Max
measured values are also recorded in addition
to the current measured value
Pre-Trigger time Define the period of time in which the mea-
sured values have already been recorded prior
to the trigger being triggered
Trigger type Select the trigger type for the measurement
Auto repeat When the option is activated, the measurement
is started again the next time the trigger is trig-
gered after automatic storage of the previous
measurement
Ring buffer If this option is activated, the defined memory
area is continuously overwritten by current
measured data
210
The Parker Service Master CONNECT V1.0/04/20