After selecting the Trigger logic measuring method, the device dis-
plays predefined parameters to complete the measurement.
1 Tap on the
button.
Ä The configuration window opens in which to define the param-
eters.
2 Select a trigger and define the parameters trigger according to
your application.
The table below contains information on the triggers available:
Designation Description
Keystroke Measurement is started by tapping on the corre-
sponding button
Level Measurement is started/stopped on exceeding or
dropping below a limit value
Window Measurement is started/stopped on exceeding or
dropping below one of two limit values for a defined
measuring range
Time Measurement is started/stopped at a defined mo-
ment of time
External Measurement is started/stopped on transition of a
digital signal from High to Low
Measurement is started/stopped on transition of a
digital signal from Low to High
Channel
warning value
Measurement is started/stopped when defined
warning values occurs
Channel alarm
value
Measurement is started/stopped when defined
alarm values occur
Duration Measurement stops after a defined time period has
expired
INFORMATION
Please note that the D-IN connection must be activated for the exter-
nal trigger. Refer to Chapter &"D-IN/D-OUT F1/F2" on Page232.