EasyManua.ls Logo

Permobil ICS - Operate Standing Sequence and Adjust Footplate Clearance; Store Final Stand Position and Settings

Permobil ICS
86 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
47
47
4. Operate the standing sequence and adjust the footplate clearance:
a) Press Button #1 to begin standing the seat.
b) Watch the footplate wheels as the seat is stood, raise the Seat Elevator using
Button #2 to prevent the footplate wheels from colliding with the ground.
c) Continue to use Button #1 to stand the seat until the Stand Angle is at the desired position.
d) Adjust the height of the Seat Elevator for proper footplate wheel clearance, using Button #2 and #6.
Proper Footplate Wheel Clearance:
If the user is in the wheelchair, the Footplate Wheels must touch the ground.
If the wheelchair is empty, there should be 10mm between the Footplate Wheels and the ground.
5. Store the final stand position and settings:
a) Press and Hold Button #8 for two seconds.
b) Release the button after a short beep sounds and LEDs #7, #8 turn green.
c) If LED #7 does not turn green, the store was not successful. The elevator is too high, lower it
and try again.
In this step, the selected Standing Sequence, desired Final Standing Angle and proper Footplate
Clearance are permanently stored by the system. This information will be used when the appropriate
Stand Button is activated and the seating system will stop at this final position.
Programming VS - Specific Attributes

Table of Contents