Draft - 1 Aug 08
ST/AR Arrhythmia Algorithm
ST/AR Arrhythmia Monitoring
7-5
ST/AR
Arrhythmia
Analysis
For information on arrhythmia detection, refer to the following documentation:
ST/AR Algorithm - Arrhythmia Monitoring Application Note,
#452298193051
IntelliVue Information Center Instructions for Use and Online Help
IntelliVue Telemetry does not have a dedicated cardiotach. Instead, the
arrhythmia cardiotach at the Information Center is used. Therefore, the ST/AR
Arrhythmia algorithm is always on for all IntelliVue Telemetry patients, and
cannot be turned off.
ST/AR provides Heart Rate and PVC Rate numerics and alarm detection for the
conditions listed in the following table. There are two detection levels: Basic and
Enhanced. Enhanced includes the Basic alarms.
Basic & Enhanced Arrhythmia Detection
Basic Arrhythmia Detection Enhanced Arrhythmia Detection
Asystole Non-Sustain VT
V-Fib/Tach Vent Rhythm
V-Tach Run PVCs
Brady yyy < xxx Pair PVCs
Tachy yyy > xxx Pause
HR yyy > xxx Missed Beat
HR yyy < xxx SVT
PVCs > xxx/min R-on-T PVCs
Pacer Not Capturing Vent Bigeminy
Pacer Not Pacing Vent Trigeminy
Multiform PVCs
Afib
Irregular HR

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Philips IntelliVue TRx and is the answer not in the manual?

Philips IntelliVue TRx Specifications

General IconGeneral
BrandPhilips
ModelIntelliVue TRx
CategoryTransceiver
LanguageEnglish

Related product manuals