Draft - 1 Aug 08
Transceiver Operation when Connected to TeleMon
Telemetry Functions at the Information Center & TeleMon
9-19
Transceiver Operation when Connected to TeleMon
When connected to TeleMon, the following differences in transceiver operation
will be in effect.
Important—The transceiver must have good batteries installed before being
connected to TeleMon.
Alarms
Alarm Suspend/Pause and Standby can be initiated only from TeleMon,
not from the transceiver or the Information Center. The alarm suspend
indicator on the transceiver will accurately reflect the current state of
alarm pause.
Note—If the transceiver is connected to TeleMon, silencing an active alarm at
TeleMon silences the alarm at TeleMon only. It has no effect on the alarms at
the Information Center.
ECG Operation
Vb, the second V-lead in 6-wire lead-set, is not supported.
After a change in leadset, the TeleMon returns to the default ECG
settings, and arrhythmia relearn occurs automatically. Be sure to check the
monitoring leads after you switch leadsets. See “Relearning” on page 6-
28.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Philips IntelliVue TRx and is the answer not in the manual?

Philips IntelliVue TRx Specifications

General IconGeneral
BrandPhilips
ModelIntelliVue TRx
CategoryTransceiver
LanguageEnglish

Related product manuals