EasyManua.ls Logo

Philips IntelliVue TRx - Cleaning the Transceiver

Philips IntelliVue TRx
274 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Draft - 1 Aug 08
Cleaning and Sterilization
Maintenance, Cleaning & Troubleshooting
11-5
Cleaning the Transceiver
Caution
Use of abrasive cleaning materials, or disinfectants or cleaning agents not listed
herein, on any part or component of the IntelliVue TRx or TRx
+
transceiver may
damage the components. See the list of unsupported cleaners in the table below.
Caution
When cleaning the Monitor/Service port, do not use any stiff, rigid instruments,
tools, or other devices to clean debris in the port, as such actions will damage the
connector pins. A forceful water stream may be used after soaking for 5 minutes
to flush the port if necessary.
Approved
Cleaners
Cleaner Active Ingredient
Isopropyl Alcohol
based
Isopropyl Alcohol (>
70%)
Ethyl Alchohol
based
Ethyl Alcohol (>
70%)
Antibacterial Soap
and Water*
Various
Hydrogen
Peroxide
Hydrogen Peroxide (3%)
Chlorine Bleach Sodium Hyprochlorite (1:10 concentration, mixed <
24 hours)

Table of Contents

Related product manuals