Draft - 1 Aug 08
Cleaning and Sterilization
Maintenance, Cleaning & Troubleshooting
11-5
Cleaning the Transceiver
Caution
Use of abrasive cleaning materials, or disinfectants or cleaning agents not listed
herein, on any part or component of the IntelliVue TRx or TRx
+
transceiver may
damage the components. See the list of unsupported cleaners in the table below.
Caution
When cleaning the Monitor/Service port, do not use any stiff, rigid instruments,
tools, or other devices to clean debris in the port, as such actions will damage the
connector pins. A forceful water stream may be used after soaking for 5 minutes
to flush the port if necessary.
Approved
Cleaners
Cleaner Active Ingredient
Isopropyl Alcohol
based
Isopropyl Alcohol (>
70%)
Ethyl Alchohol
based
Ethyl Alcohol (>
70%)
Antibacterial Soap
and Water*
Various
Hydrogen
Peroxide
Hydrogen Peroxide (3%)
Chlorine Bleach Sodium Hyprochlorite (1:10 concentration, mixed <
24 hours)