Draft - 1 Aug 08
Telemetry Functions at the Information Center
9-6 Telemetry Functions at the Information Center & TeleMon
Patient
Configurable
Settings in
Telemetry
Setup
The Telemetry Setup window enables you to configure the transceiver for
patient-specific settings. All patient-specific settings will be reset to the unit
defaults upon patient discharge. To access the window, from the Patient
Window click Telemetry Setup.
The following settings can be adjusted in this window.
Patient-Configurable Settings in Telemetry Setup
Control Function Setting Choices
Factory
Default
Telemetry
Button
Determine the Information Center
response when Telemetry Button
is pressed.
Nurse Call - generate nurse
call alarm
Record - generate a
recording strip
Nurse Call and Record -
generate nurse call alarm and
recording strip
None
Nurse Call
Telemetry
Device:
Volume
Set the volume level for all
adjustable sounds on the
transceiver.
1 (low), 2, 3, 4, 5 (high) 3
Telemetry
Device:
Mute
Enable/disable Spot Check and
pulse detection tones on the
transceiver.
enable (mute)
disable
enable
Fixed Pacer
Amplitude
Sets the appearance of the pacer
spikes to a fixed size as they
appear in the patient window.
enable
disable
disable

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Philips IntelliVue TRx and is the answer not in the manual?

Philips IntelliVue TRx Specifications

General IconGeneral
BrandPhilips
ModelIntelliVue TRx
CategoryTransceiver
LanguageEnglish

Related product manuals