Measurement Specifications
12-18 Safety Standards & Specifications
SpO
2
Parameter Specification
SpO
2
Measurement
Range (Calibration and
Display)
0 to 100%
SpO
2
Accuracy See table following.
SpO
2
Resolution 1%
SpO
2
Numerics -
Averaging
10 seconds
Note—The update rate for the SpO
2
pulse
oximetry value and pulse rate is typically 1
second. This can be extended to a max. 60 s
when NIBP is measured on the same limb, with a
corresponding INOP message after a max. of 30
s, indicating that the displayed values are not
current values.
The effect of SpO
2
pulse oximetry on data
averaging is internally controllable by the
transmitter, with no user controls.
SpO
2
& Pulse Numerics
- Update Rate
Transmitted once per second.
Pleth Wave- Sampling
Rate
125 sps
Technical Alarms
(INOPs)
Triggered if the sensor is disconnected, if a pulse
is not detected, if the signal is noisy, if light
interference is detected, if the sensor is defective,
if the measurement is erratic, or if the module is
malfunctioning

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Philips IntelliVue TRx and is the answer not in the manual?

Philips IntelliVue TRx Specifications

General IconGeneral
BrandPhilips
ModelIntelliVue TRx
CategoryTransceiver
LanguageEnglish

Related product manuals