SpO
2
Safety Information
SpO
2
Monitoring 8-3
WarningWarning
Interference leading to inaccurate measurements can be caused by:
- High levels of ambient light (Hint: cover application site with opaque
material)
- Electromagnetic interference
- Excessive patient movement and vibration.
WarningWarning
This equipment is not suitable for use in the presence of a flammable
anesthetic mixture or oxygen concentrations greater than 25% (or partial
pressures greater than 27,5 kPa /206.27 mmHg).
WarningWarning
Disposable SpO
2
sensors can be damaged and lead to patient harm if they
become wet. Wet sensors must be replaced immediately.
SpO
2
Information
for the User
The pulse oximeter is calibrated to indicate functional oxygen saturation
(fractional oxyhemoglobin), and displayed results can range from 0 to 100%.
A 10 second averaging filter is used in the calculation of the result. Displayed
results are typically updated every second, but the update period can be
automatically delayed by up to 30 seconds in the presence of noise.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Philips IntelliVue TRx and is the answer not in the manual?

Philips IntelliVue TRx Specifications

General IconGeneral
BrandPhilips
ModelIntelliVue TRx
CategoryTransceiver
LanguageEnglish

Related product manuals