EasyManua.ls Logo

Philips IntelliVue TRx - Page 156

Philips IntelliVue TRx
274 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Draft - 1 Aug 08
SpO
2
Safety Information
8-4 SpO
2
Monitoring
Physiological SpO
2
alarm signals will be generated at the Information Center.
The SpO
2
low limit can be set between 50 and 99% inclusive, in 1%
increments. The SpO
2
high alarm limit can be set between 51 and 100%
inclusive, in 1% increments. The maximum delay between the physiological
alarm condition and alarm signal generation at the central station is 10 seconds.
Pulse rate is also derived from the pulsatile SpO
2
measurement, and displayed
results can range from 30 to 300 bpm. There is no alarm function for pulse rate.
The pleth wave is auto-scaled to maximum display size. It decreases only when
the signal quality becomes marginal. Pleth wave size is NOT directly
proportional to the pulse volume.

Table of Contents

Related product manuals