Draft - 1 Aug 08
Pairing Non-networked Devices
10-20 Pairing Monitoring Devices
The following table summarizes alarm behavior when a transceiver is paired
with an IntelliVue bedside monitor without network connection. For detailed
information, see the IntelliVue Patient Monitor Instructions for Use and the
IntelliVue Information Center Instructions for Use.
Where Initiated Effect at Bedside Effect at Information Center
Silence
Bedside Silences bedside alarms. No effect.
Information Center No effect Silences telemetry alarms.
Suspend/Pause
Bedside Suspends/Pauses bedside and
telemetry alarms.
Suspends/Pauses telemetry alarms.
All Arrhythmia alarms are off.
Information Center No effect
Suspend/Pause not available at the
Information Center
No effect.
Suspend/Pause not available at the
Information Center
Standby
Bedside Bedside put into standby state. Measurements from bedside
disappear. Telemetry monitoring
continues.
Information Center N/A
Standby not available at the
Information Center.
N/A
Standby not available at the
Information Center.
Resume Monitoring
Bedside Bedside resumes monitoring Measurements from the bedside
reappear, for example NBP and/or
Temp, in the Patient Window.
Information Center N/A
Resume Monitoring not available at
the Information Center.
N/A
Resume Monitoring not available at
the Information Center.