Draft - 1 Aug 08
SpO
2
Product Safety 2-15
WarningWarning
If you measure SpO
2
on a limb that has an inflated NBP cuff, a non-
pulsatile SpO
2
technical alarm can occur. If the monitor is configured to
suppress this alarm, there can be a delay of up to 60 seconds in indicating
critical patient status, such as sudden pulse loss or hypoxia.
WarningWarning
Removal of the SpO
2
sensor during Continuous SpO
2
monitoring results in
a "No Sensor" technical alarm. Silencing this technical alarm turns the
SpO
2
measurement off. There is no technical alarm for a “No Sensor”
condition in Spot Check mode.
WarningWarning
This equipment is not suitable for use in the presence of a flammable
anesthetic mixture or oxygen concentrations greater than 25% (or partial
pressures greater than 27,5 kPa /206.27 mmHg).