Turning the Transceiver On/Off
Basic Operation
4-5
Resuming
Monitoring
4 Select the appropriate location, then select
Suspend Monitoring.
The message “Telemetry Standby” and location, if selected, are
displayed in the sector.
Note—If the transceiver is paired with an IntelliVue Patient
Monitor, monitoring will be suspended at both the transceiver and
the IntelliVue Patient Monitor as well as at the Information Center.
The messages "Monitor Standby" and "Telemetry Standby" will be
displayed in the patient sector. If paired directly with an IntelliVue
MP5 Patient Monitor, the message "Monitor Standby" only will be
displayed.
Step Action
Step Action
1 If the standby period has not expired:
•Press the Check button to re-establish contact with the
Information Center. The single beep at the transceiver
verifies that monitoring has resumed.
If you hear a double beep, click Resume Monitoring at the
Information Center. If there’s still no association, contact the
service provider.
If the standby period has expired:
Monitoring resumes automatically if the patient is in the
coverage area. Press the Check button to verify the
resumption of monitoring. You should hear a single beep.
If a new patient is connected:
•Press the
Check button, or select the Resume Monitoring in
the appropriate patient sector.
Important—When an EASI patient comes out of Standby, the lead
settings are reset to the default EASI leads.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Philips IntelliVue TRx and is the answer not in the manual?

Philips IntelliVue TRx Specifications

General IconGeneral
BrandPhilips
ModelIntelliVue TRx
CategoryTransceiver
LanguageEnglish

Related product manuals