Measuring ECG
ECG Monitoring
6-5
The transceiver detects the inserted leadset type and automatically determines
the ECG measurement and transmitted leads. The Leadset Insertion Guide on
the device will assist you in ensuring the correct measurement during transceiver
usage. If the transceiver senses the insertion of a leadset not matching the
current configuration, the transceiver will source the ECG per the leadset
detected and issue and auditory signal to indicate the condition. An INVALID
LEADSET INOP is generated at the Information Center.
Note—The labels and colors of the ECG electrodes differ according to the
standards that apply for your hospital. The electrode placement references and
illustrations in this chapter use the AAMI labels and colors. See the table below
for additional label and color information.
Electrode Labels Electrode Colors
AAMI EASI IEC AAMI IEC
RA I R White Red
LA S L Black Yellow
LL A F Red Green
RL N N Green Black
V/Va E C/Ca Brown White
Vb Cb Brown/White White/Red
V2 C2 Brown/Yellow White/Yellow
V3 C3 Brown/Green White/Green
V4 C4 Brown/Blue White/Brown
V5 C5 Brown/Orange White/Black
V6 C6 Brown/Violet White/Violet