What to do next
The duration of the cup filling can also be programmed as follows:
1. Place an empty cup to the cup holder.
2. Press Program.
3.
Press Cup fill / bowl rinse twice. The text ‘C n.n’ will appear on the
control panel.
4. Push the foot control pedal down. The cup filling time will first be set to 2
seconds and will start to increase after pushing the foot control pedal for
2 seconds. The pedal can be released and then pushed again, and the
time continues to increase. Continue until the cup is filled to the desired
level.
5.
Press Program.
22.5 Planmeca Solanna and Planmeca Solanna Vision operating lights
NOTE
The Planmeca Solanna Vision operating light is not available for the
Planmeca Compact i dental unit.
22.5.1 Intensity
About this task
NOTE
The light intensity can also be adjusted from the operating light, see sections
"Adjusting intensity of operating light" on page 72 and "Adjusting intensity of
operating light in composite mode" on page 73.
Steps
1. Press Program.
2. Press Operating light.
22 Programming
User's manual Planmeca Compact i 103

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Planmeca Compact i Classic v2 and is the answer not in the manual?

Planmeca Compact i Classic v2 Specifications

General IconGeneral
BrandPlanmeca
ModelCompact i Classic v2
CategoryDental equipment
LanguageEnglish

Related product manuals