8. Start the short flushing cycle.
Balanced instrument arms: Start the flushing cycle by bending the
instrument arms of those instruments (including syringe) that are to be
flushed to an angle of at least 90° and simultaneously pressing
Instrument spray (you will hear a signal tone) and holding it for 3
seconds. Release the button and arms when you hear a second signal
tone.
Hanging-tube instruments: Start the flushing cycle by pressing
Instrument spray (you will hear a signal tone) and holding it for 3
seconds. Release the button when you hear a second signal tone.
The system will first identify the instruments and then each instrument
hose is flushed with air and water for 30 seconds in its turn.
9. Turn the unit off and let the disinfectant affect in the unit overnight.
NOTE
Leave the instruments in the flushing holder overnight.
25.2.2 In the morning
Steps
1. Open the dental unit door.