Note that the flushing holder cover can not be autoclaved. Remove the cover
before autoclaving the instrument flushing holder.
24.4.3 Clean-water bottle
Use a bottle brush and a mild soap- and water solution to clean the clean-
water bottle once a week.
To remove the clean-water bottle from the dental unit, first turn the Water
bottle switch toward 'OFF'. Then, take a hold of the bottom of the bottle, turn
the bottle slightly so that the pin moves along the groove and pull the bottle
downwards.
24.5 Suction system
NOTE
Dry the suction tube holders and the tube bushings properly after cleaning.
Wet surfaces might disturb suction tube recognition.
24.5.1 In the morning
Steps
1. Rinse each suction tube with 0.5 litres of water by slowly aspirating water
and air through the suction handpieces.
2. Wipe the suction handpieces with Planmeca approved surface
disinfectant.