4. Let the solution affect overnight. Rinse the suction system the next
morning with 2 l of water.
5. Rinse the bowl with water the next morning by pressing Cup fill / bowl
rinse.
24.5.5 Cleaning suction handpieces
The following explains how the suction handpieces can be disassembled for
cleaning.
NOTE
We recommend that the suction handpieces be replaced once a year.
High-volume suction handpiece
Unscrew the handpiece from the suction tube for cleaning.
If necessary, the handpiece can be completely disassembled for more
thorough cleaning. The handpiece parts can be disinfected in a washer-
disinfector at 93°C, then optionally autoclaved at 134°C.
24 Cleaning and disinfection
User's manual Planmeca Compact i 127