Remove the supplementary holder by pulling it out from the Flexy-holder. To
replace it, push it firmly into its place.
Remove the holder for the intraoral scanner by pressing the trigger (see
arrow in picture) and pulling the holder out from the Flexy-holder. To replace
the holder for the intraoral scanner, push it firmly into its place.
11.3 Removing and replacing suction tubes
Steps
1. To minimise contamination risk, perform suction cleaning.
For instructions, see section "Suction cleaning" on page 115.
2. Remove the side cover plate by pulling it out from the cuspidor.
3. Detach the securing ring (1) from the suction tube.
NOTE
You can grease the securing ring with non-toxic vaseline to make it
easier to remove.