EasyManuals Logo

Planmeca Compact i Classic v2 User Manual

Planmeca Compact i Classic v2
190 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #132 background imageLoading...
Page #132 background image
24.5.2 After each patient
Steps
1. Remove the used aspirating tips.
2. For hygienic and operational reasons, empty one glass of water (100 -
200 ml) with each suction tube by aspirating water and air through the
suction handpiece. Do this even if only the saliva suction tube has been
used.
3. Wipe the suction handpieces with Planmeca approved surface
disinfectant.
4. Wipe the suction holder / Flexy-holder, including the tablet holder, with
Planmeca approved surface disinfectant.
Results
NOTE
Do not use a spray disinfection solution for the suction arms and holders.
NOTE
Clean the tablet according to the instructions given by the manufacturer.
24.5.3 After each working day
Steps
1. Disinfect the suction system either by running the suction cleaning
program or by using OroCup. For information on the suction cleaning
program, see section "Suction cleaning" on page 115.
2. When the suction system has been rinsed, disinfect the suction
handpieces in a washer-disinfector at 93°C, then optionally autoclave
them at 134°C.
3. Flexy-holder: Disinfect the supplementary holders, suction tube holders
(including rollers) and instrument holder in a washer-disinfector at 93°C.
4. Flexy-holder: Wipe the holder for the intraoral scanner with Planmeca
approved surface disinfectant.
24 Cleaning and disinfection
124 Planmeca Compact i User's manual

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Planmeca Compact i Classic v2 and is the answer not in the manual?

Planmeca Compact i Classic v2 Specifications

General IconGeneral
BrandPlanmeca
ModelCompact i Classic v2
CategoryDental equipment
LanguageEnglish

Related product manuals