Help messages
HELP
CODE
MEANING ACTION OR EXPLANATION
H 90.2 ActiveAqua prefilter expired. Switch off the dental unit. Replace the
prefilter. After switching on the dental unit
press the Reset prefilter button to confirm
the prefilter replacement.
To temporarily ignore this notification, select
any instrument or press Cancel.
H 90.3 ActiveAqua water container is empty. Water container fill is slow or the low level
sensor does not sense water.
H 90.4 ActiveAqua water container overflow. Check ActiveAqua level sensors and water
container.
H 90.5 ActiveAqua annual maintenance overdue. Annual ActiveAqua maintenance must be
performed. Contact service.
H 90.6 Measured chlorine level low. Normal use chlorine level increased by
+15%.
H 90.7 Measured chlorine level OK. No change to normal use chlorine level.
H 90.8 Measured chlorine level high. Normal use chlorine level decreased by
-5%.
H 91 Stop (Help) button pressed.
H 92 Help (Stop) button pressed.
H 93 This button has no function at the moment
because there is no motorised light.
H 94 Clock calibration failed. Maximum
calibration range exceeded.
H 95 Patient chair automatic position cannot be
reached because legrest is unlocked or
armrests are blocking movement.
H 96 PlanID tag already in use. PlanID tag assigned to another user.
H 96.1 Different user already logged in. The unit has rejected a login request
because a different user is already logged
in.
H 97 Failed to assign PlanID tag to user. Check Romexis connection.
H 98 User not found. User not found with this PlanID tag. Create
a new user or select an existing Romexis
user and assign the PlanID tag.
H 98.1 User login failed. Retry login or reboot the unit. If problem
persists, contact service.
H 98.2 RFID unknown. The RFID is unknown to the system.
H 99 Unit has been switched off without
performing long flushing after water has
been used. Perform long flushing.
This help message can be disabled, contact
your Planmeca dealer.
HE200 Detected new software update. Confirm to
install?
HE201 Detected new software update SD-card.
Confirm to install 5.3.0.9.R ?
27 Help and error messages
148 Planmeca Compact i User's manual

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Planmeca Compact i Classic v2 and is the answer not in the manual?

Planmeca Compact i Classic v2 Specifications

General IconGeneral
BrandPlanmeca
ModelCompact i Classic v2
CategoryDental equipment
LanguageEnglish

Related product manuals