18
 
6.3. Return to weighing 
 
Introduced changes in settings will be saved after return to weighing 
mode with procedure of saving changes. 
 
 
 
Fig. 6. Return to weighing 
 
After introducing all changes in parameters settings press several times  ESC 
key. When display indicates message, choose one of two options: 
ENTER – save changes 
ESC – without saving  
 
 
 
7. WEIGHING  
 
Basic conditions which should be fulfilled to ensure reliability of measurements: 
9  stable base for balance, 
9  choice of appropriate parameters for existing external conditions  
 
Before measurements or in case of essential changes of external conditions, 
balance should be calibrated according to manual (for verified balances, 
function of calibration is not accessible for user)  
Before measurements check if unloaded balance indicates „precise zero” – 
indicated pictogram -  
  - in left lower corner (only when P4  06 Autozero 
parameter is set on 1: yes or P4 07 for second platform), if balance doesn’t 
indicates zero press key: 
. 
 
If conditions for zeroing are unfavourable (lack of stable result), display 
indicates horizontal lines. After exceeding set zeroing time of balance will 
return to weighing without performing zeroing. In such case wait for 
stabilization of conditions and press again 
 key. 
 
Place on the pan weighed load and when results are stable (indicated 
) 
read on the display result of weighing.