26
 
11. SETTING WORK PARAMETERS 
 
User by means of appropriate setting of parameters from menu group <P4 
Readout> can adjust the balance to existing operating conditions (filter) and 
one’s expectations (refreshing, autozero, last digit displaying)  
 
 
 
Fig. 13. Submenu Readout – internal settings 
 
11.1. Filter setting 
 
For perfect conditions filter can be set as very fast, however if conditions are 
bad (vibrations, draughts) filter should be set as slow or as very slow. 
Effectiveness of filter operating is different for weighing range. Filter works 
slower in time of “reaching” the final result, however stronger when mass gets 
to set range of filter operating (parameter range of filter operating available only 
in service menu – user do not have access to it). 
 
 
11.2. Median filter setting 
 
Task of median filter is to eliminate single big disturbances. By setting digital 
value determines speed of filter operating. For zero option filter operating is off.