3. GETTING STARTED 
 
3.1 Conditions of appropriate use 
 
9  place balance on a stable and flat table or free from vibrations 
9  balance should not be exposed to draughts and sudden air movements 
9  balance should be placed in room of stable temperature and humidity 
9  balance should be placed far from heat sources 
9  temperature of the room should be +15°C ÷ +30°C (for WLT balances)  
9  if static electricity influence balances, indications, its base should be 
earthed. Earthling clamp is in the back part of balance base 
9  balances should be levelled according to a level condition indicator to 
guarantee the appropriate weighing accuracy 
 
 
3.2 Time of warming-up 
 
Before performing measurements wait until balance reaches temperature 
stabilization. It is so-called warm-up time. It takes about 15 minutes.  
For weighing instruments that have been stored in low temperatures (e.g. 
during winter) the warm-up time is about 2 hours. 
 
The indications of display may change during temperature stabilization.