78
 
26.4.2. Continuous transmission 
 
Results are transmitted in basic units. Transmission can be disabled/enabled 
using RS232 command: 
 
C1 CR LF – Switch on continuous transmission in basic unit 
C0 CR LF – Switch off continuous transmission in basic unit 
 
or using the parameter: settings/ RS-232/Aut. printout/continuous 
 
The frame length is constant (21 characters). 
 
Frame format: 
 
[SI] [space] [stability] [space] [sign] [mass] [space] [unit] [CR LF] 
 
Stability character – [space] if stable 
[?] if not stable 
[^] if an indication over the range 
[v] if an indication below the range 
sign – [space] for positive values or [-] for negative values 
mass – 9 characters justified to the right 
unit – 3 characters justified to the left 
 
 
26.4.3. Continuous transmission „with pauses” 
 
Contin
uou
s transmission „with pauses” allows users to determine the time 
interval between the subsequent frames in the parameter interval. 
His type of printouts can be enabled only in parameters: Settings/RS-232/ Aut. 
printouts/with interv. 
 
The frame format is the same as in basic simple transmission. 
The interval is set in: Setting/RS-232/Interval 
 
 
26.4.4. Stable result transmission 
 
The scale a
u
tomatically generates after each stable result separated by a 
stable result below the Min. mass that can be set in parameters. 
 
The intended use of this printout is for slip printers so it doesn’t have a starting 
characters and the length can vary from 18 to 20 characters.