ppbRAE 3000 User’s Guide
13
IEC 60079-11:2006
(IIC: 059-3051-000 Li-ion bat pack
or 059-3054-000 NiMH bat pack;
IIB: 059-3052-000 alkaline bat pack)
EM Interference: Highly resistant to EMI/RFI. Compliant with
EMC R&TTE (RF Modules)
Alarm Setting: Separate alarm limit settings for Low, High,
STEL and TWA alarm
Operating Mode: Hygiene or Search mode
Alarm: Buzzer (95dB at 30cm) and flashing red
LEDs to indicate exceeded preset limits, low
battery voltage, or sensor failure
Alarm Type: Latching or automatic reset
Real-time Clock: Automatic date and time stamps on
datalogged information
Datalogging: 260,000 points with time stamp, serial
number, user ID, site ID, etc.
Communication: Upload data to PC and download instrument
setup from PC via USB on charging station.
Wireless Network: Mesh RAE Systems Dedicated Wireless
Network (or WiFi network for WiFi-
equipped instruments)
Sampling Pump: Internally integrated. Flow rate: 450 to 550
cc/min.
Wireless Frequency: ISM license-free band, 902 to 907.5 MHz
and 915 to 928 MHz, FCC Part 15, CE
R&TTE,
IEEE 802.11 b/g bands (2.4 GHz)
Modulation: 802.15.4 DSSS BPSK
RF Power (Tx): 10dBm
Temperature: -20º C to 50º C (-4º to 122º F)
Humidity: 0% to 95% relative humidity (non-
condensing)