ppbRAE 3000 User’s Guide
39
Optional Accessories
Calibration Adapter
The calibration adapter for the instrument is a simple 6-inch Tygon
tubing with a metal adapter on one end. During calibration, simply
insert the metal adapter into the regular gas inlet probe of the
instrument and the tubing to the gas regulator on the gas bottle.
Calibration Regulator
The Calibration Regulator is used in the calibration process. It
regulates the gas flow rate from the Span gas cylinder into the gas
inlet of the instrument during calibration process. The maximum flow
rate allowed by the flow controller is about 0.5L/min (500 cc per
min.). Alternatively, a demand-flow regulator or a Tedlar gas bag
may be used to match the pump flow precisely.
Organic Vapor Zeroing Kit
The Organic Vapor Zeroing Kit is used for filtering organic air
contaminants that may affect the zero calibration reading. To use the
Organic Vapor Zeroing Kit, simply connect the filter to the inlet port
of the instrument.
AutoRAE 2 Automatic Test & Calibration
System
The AutoRAE 2 Automatic Test and Calibration System for RAE
Systems portable gas monitors makes compliance with monitor test
and calibration requirements as easy as pressing a button. Simply
cradle the monitor and the system will take care of all calibration,
testing, and recharging.
The AutoRAE 2 is a flexible, modular system that can be configured
to meet your calibration requirements effectively and efficiently. An
AutoRAE 2 system can be as simple as a single cradle deployed in
standalone mode to calibrate one instrument at a time, or as powerful
as a networked, controller-based system supporting ten monitors and
five distinct calibration gas cylinders.