ppbRAE 3000 User’s Guide
106
Inlet flow too low
Reasons: Pump diaphragm damaged
or has debris.
Flow path leaks.
Solutions: Check flow path for leaks;
sensor module O-ring, tube
connectors, Teflon tube
compression fitting.
Call Technical Support at
+1 408-752-0723
or toll-free at
+1 888-723-4800
“Lamp” message
during operation
Reasons: Lamp drive circuit.
Weak or defective PID
lamp, defective.
Solutions: Turn the unit off and back
on.
Replace UV lamp