ppbRAE 3000 User’s Guide
21
Operating The Instrument
The instrument is designed as a broadband VOC gas monitor and
datalogger for work in hazardous environments. It gives real-time
measurements and activates alarm signals whenever the exposure
exceeds preset limits. Prior to factory shipment, the instrument is
preset with default alarm limits and the sensor is pre-calibrated with
standard calibration gas. However, you should test the instrument and
verify the calibration before the first use. After the instrument is fully
charged and calibrated, it is ready for immediate operation.
Turning The Instrument On
1. With the instrument turned off, press and hold [MODE].
2. When the display turns on, release the [MODE] key.
The RAE Systems logo should appear first. (If the logo does not
appear, there is likely a problem and you should contact your
distributor or RAE Systems Technical Support.) The instrument is
now operating and performs self tests. If any tests (including sensor
and memory tests fail), refer to the Troubleshooting section of this
guide.
Once the startup procedure is complete, the instrument shows a
numerical reading screen with icons. This indicates that the
instrument is fully functional and ready to use.