Duty Cycle
The pump’s duty cycle is the ratio of its on time to off time. The duty
cycle ranges from 50% to 100% (always on), and the period is 10
seconds. Therefore, a duty cycle of 60% means that the pump is on
for 6 seconds and off for four seconds. Duty cycling is employed by
the instrument to clean the PID. A lower duty cycle has a greater
effect on keeping the PID clean than a higher duty cycle.
Important! Pump duty cycling is interrupted when the instrument
senses a gas. The pump’s duty cycle is disabled when the
measurement is greater than the 2ppm threshold and is re-enabled
when the reading falls below 90% of the threshold (1.8 ppm).
1. Press [Y/+] to increase the value.
2. When you have completed your selection, press [MODE].
Press [Y/+] to save the new duty cycle value.
Press [N/-] to undo the change and move to the next sub-
menu.