Response time (T
90
): 2 seconds
Accuracy 10 to 2000 ppm: 3% at calibration point.
(Isobutylene):
PID Detector: Easy access to lamp and sensor for cleaning
and replacement
Correction Factors: Over 200 VOC gases built in (based on RAE
Systems Technical Note TN-106)
Calibration: Two-point field calibration of zero and
standard reference gases
Calibration Reference: Store up to 8 sets of calibration data, alarm
limits and span values
Inlet Probe: Flexible 5" tubing
Radio module: Bluetooth (2.4GHz) or RF module (433MHz,
868MHz , 915MHz, or 2.4GHz)
Keypad: 1 operation key and 2 programming keys; 1
flashlight switch
Direct Readout: Instantaneous, average, STEL, TWA and
peak value, and battery voltage
Intrinsic Safety: US and Canada: Class I, Division 1, Groups
A, B, C, D
Europe: ATEX (0575 Ex II 2G Ex ia
IIC/IIB T4 Gb)
KEMA 07 ATEX 0127
Complies with EN60079-0:2009,
EN60079-11:2007
IECEx CSA 10.0005 Ex ia IIC/IIB T4 Gb
Complies with IEC 60079-0:2007,