EB200 Manual DISPlay subsystem
4052.2000.02 4.30 E-7
. :ENABle <Boolean>
Enable or disable the manual control and display (display, rollkey and keys).
CONTROLLED BY REMOTE is displayed on the front panel in the "disable" state. The string can be
configured by means of the DISPLAY:ENABLe:LABel command. This command is advantageous if,
in systems, EB200 is exclusively controlled by Remote.
When the front-panel controls are switched off, additional computing power becomes available in the
processor and time-critical operations (scans, transmission of large volumes of data) run much more
quickly.
Parameters:
ON enable manual control
OFF disable manual control
*RST state:
ON
Example:
DISPlay:ENABle ON
. :ENABle?
Query of state of manual control .
Parameters:
none
Result:
1 manual control enabled
0 manual control disabled
Example:
DISPlay:ENABle? -> 1