EB200 Manual ROUTe subsystem
4052.2000.02 4.55 E-7
. OPEN
. . :ALL
Select no antenna (antenna number 0 is set)
Parameters:
none
*RST state:
none since command is an event
Example:
ROUTe:OPEN:ALL
. :PATH
. . :CATalog?
Request of a list with currently defined antenna names
If the option EB200FS (Field Strength Measurement) is fitted, specific names are assigned to antenna
numbers 1 to 5 (see example below).
Parameters:
none
Result:
One string for each antenna name, separated by comma
<name_string1>,<name_string2>,...
If no antenna name is defined, a zero string ("") is output.
Example:
ROUTe:PATH:CATalog? -> "HE200 HF ", "HE200 20-200M", "HE200 .2-0.5G",
"HE200 .5-3GHZ", "GENERIC DIPOL", "ANT06", "ANT07",.....
. . [:DEFine] <path_name>, <channel_list>
Assignment of an antenna name to the antenna number
Parameters:
<path_name> This is a <Character Data> with up to 13 characters which
contains the antenna name. The path name may also be put in
inverted commas to allow the use of blanks.
<channel_list> see ROUTe:CLOSe
*RST state:
All names are maintained also after *RST.
Example:
ROUTe:PATH HE110, (@10)