Maintenance
R&S
®
TS7124M
152User Manual 1525.5394.02 ─ 05
8 Maintenance
Familiarize yourself with the residual risks and potentially dangerous situations.
The chamber has no factory default settings.
● Compliant maintenance........................................................................................ 152
● Maintenance intervals........................................................................................... 152
● Checking............................................................................................................... 153
● Cleaning................................................................................................................155
● Storage..................................................................................................................155
● Exchanging the trailing cable in the energy chain.................................................156
● Contacting customer support................................................................................ 161
8.1 Compliant maintenance
To retain the functional readiness and long operational life of the RF shielded box, per-
form maintenance tasks according to the intervals specified in Chapter 8.2, "Mainte-
nance intervals", on page 152.
Operating the RF shielded box involves producing some inevitable waste and expend-
ables. Handle these in an environmentally friendly way.
8.2 Maintenance intervals
The following table shows the maintenance tasks that are to be carried out in the sta-
ted intervals on the RF shielded box. The intervals are valid for a single work shift cor-
responding to about 160 hours per month.
Table 8-1: Scheduled maintenance
Maintenance Interval Maintenance Tasks
Daily Functional checking
• Check the door's gasket for soiling, damage and wear
• Check the drawer's manual close mechanism for proper function
• Check antennas, cables and connectors for proper function
• Check the door closure for the right gap size
Weekly Absorber and shielding effectiveness checking
• Check the absorber material on the inside of the drawer and
around the chamber's door opening for damage or wear
• Check the tension force at the top cover
Every 100 000 cycles Gasket cleaning
• Clean the door's gasket every 100 000 cycles to maintain RF
shielding