Maintenance
R&S
®
TS7124M
152User Manual 1525.5394.02 ─ 05
8 Maintenance
Familiarize yourself with the residual risks and potentially dangerous situations.
The chamber has no factory default settings.
Compliant maintenance........................................................................................ 152
Maintenance intervals........................................................................................... 152
Checking............................................................................................................... 153
Cleaning................................................................................................................155
Storage..................................................................................................................155
Exchanging the trailing cable in the energy chain.................................................156
Contacting customer support................................................................................ 161
8.1 Compliant maintenance
To retain the functional readiness and long operational life of the RF shielded box, per-
form maintenance tasks according to the intervals specified in Chapter 8.2, "Mainte-
nance intervals", on page 152.
Operating the RF shielded box involves producing some inevitable waste and expend-
ables. Handle these in an environmentally friendly way.
8.2 Maintenance intervals
The following table shows the maintenance tasks that are to be carried out in the sta-
ted intervals on the RF shielded box. The intervals are valid for a single work shift cor-
responding to about 160 hours per month.
Table 8-1: Scheduled maintenance
Maintenance Interval Maintenance Tasks
Daily Functional checking
Check the door's gasket for soiling, damage and wear
Check the drawer's manual close mechanism for proper function
Check antennas, cables and connectors for proper function
Check the door closure for the right gap size
Weekly Absorber and shielding effectiveness checking
Check the absorber material on the inside of the drawer and
around the chamber's door opening for damage or wear
Check the tension force at the top cover
Every 100 000 cycles Gasket cleaning
Clean the door's gasket every 100 000 cycles to maintain RF
shielding