Installation and commissioning
R&S
®
TS7124M
22User Manual 1525.5394.02 ─ 05
5 Installation and commissioning
Familiarize yourself with residual risks and potentially dangerous situations.
See Chapter 1.1, "Safety instructions", on page 7.
Execute these activities in the same order as given in this handbook:
Choosing the operating site.................................................................................... 22
Unpacking............................................................................................................... 23
Mounting the chamber............................................................................................ 24
Grounding the RF shielded box.............................................................................. 28
5.1 Choosing the operating site
Only use the chamber indoors. The chamber casing is not waterproof.
Select an operating site that provides safe conditions for mounting and operating the
chamber.
Ensure the following:
The room has an even floor and sufficient bearing capacity.
The operating site leaves sufficient room to open the door without obstruction and
to access:
Chamber, especially the area behind the open door
Connectors on the front and rear side
Mounting brackets
The environmental conditions such as ambient temperature and humidity match
the values in the specifications.
The operating site is at an altitude of maximum 2000 m above sea level.
The environment has pollution degree 2, where only nonconductive contamination
occurs. Occasionally, temporary conductivity caused by condensation is to be
expected.
The electromagnetic compatibility (EMC) class of the chamber is class A.
Electromagnetic compatibility classes
The EMC class indicates where you can operate the chamber.
Class B equipment is suitable for use in:
Residential environments
Environments that are directly connected to a low-voltage supply network that
supplies residential buildings
Class A equipment is intended for use in industrial environments. It can cause
radio disturbances in residential environments due to possible conducted and radi-
ated disturbances. It is therefore not suitable for class B environments.