Configuration
R&S
®
TS7124M
84User Manual 1525.5394.02 ─ 05
● Rotate clockwise for a lower position of the quick clamp, leading to a stronger
tension force.
● Rotate anti-clockwise for a higher position of the quick clamp, leading to a
weaker tension force.
4. In case of anti-clockwise rotation (for a higher position), make sure to shift the
quick clamp upwards to keep it in contact with the adjustment screw.
5. Tighten the two fixing screws (2) by a Torx 9 screwdriver.
6. Check the tension force and the good operability of the quick clamp by closing it as
shown in Figure 7-8.
In the best position of the quick clamp, it is possible to close it by hand with mild
force, while the top cover is being pulled downwards by a fraction of a millimeter.
To detect this movement, hold a fingertip across the small gap between the top
cover and the chamber's body. This should only be done once the top cover is
close enough to the chamber's body (<1 mm), to avoid any risk of contusion. By
touching the gap while closing the clamp, the slight downwards movement of the
top cover can just be felt. This is a good indication for an appropriate adjustment of
the tension force.
As long as the tension force and the operability of the quick clamps are not ideal,
repeat the procedure:
● If the force is too weak (hence the top cover is not noticeably pulled down),
rotate the adjustment screw half a turn clockwise in the repeated procedure.
● If the force is too strong (hence closing the quick clamps by hand is hardly pos-
sible), rotate the adjustment screw half a turn anti-clockwise in the repeated
procedure.
7.5 Mounting a DUT holder tray
The R&S TS7124M can optionally be equipped with the DUT holder tray R&S TS-
F24P1 to be mounted to the inside of the drawer door. You can mount the tray either in
a lower or in a higher position, depending on your intended application.
Mounting a DUT holder tray