51
eRev. 1.08/15/2023
2. Replace the gasket/seal pad between the Detector and Flow Cell every time a Flow Cell is loaded.
3. Replace the gasket/seal pad between the Flow Cell and Head every time a Flow Cell is loaded.
Perform a Transmission Check (Cary 60)
Perform a transmission check after each time a Flow Cell is loaded.
Clean the Flow Cell
Clean the Flow Cell by running 0.5 M NaOH for 30 minutes after a process run.
Check Connections
Ensure all of the connections to the FlowVPX Head are secure and fastened appropriately.
Protect Optical Surfaces
Protect the optical surfaces when the system is not in use. Flow Cells can be returned to their shipment cartons, the Delivery Fiber
should be capped and bagged, and the Detector can be stored in the Detector Validation Adapter with the Transmission Tool
attached.