RIGOL Chapter 2 Command System
2-24 DL3000 Programming Guide
[:SOURce]:CURRent:VON
Syntax [:SOURce]:CURRent:VON {<value>|MINimum|MAXimum|DEFault}
[:SOURce]:CURRent:VON? [MINimum|MAXimum|DEFault]
Description Sets the starting voltage in CC mode.
Queries the starting voltage set in CC mode.
Remarks
When the input voltage increases and reaches a value above the set starting voltage
(Von), the load starts to sink the current; when the input voltage decreases and
reaches a value below the set starting voltage, the load stops sinking. The unit is V.
In the setting command, the parameter "MINimum" indicates setting the minimum
Von value; the parameter "MAXimum" indicates setting the maximum Von; and the
parameter "DEFault" indicates setting the default Von value.
In the query command, the parameter "MINimum" indicates querying the minimum
Von value; the parameter "MAXimum" indicates querying the maximum Von value;
and the parameter "DEFault" indicates querying the default Von value.
Return
The query returns a real number.
Example :SOUR:CURR:VON 5 /*Sets the starting voltage (Von) in CC mode to be 5 V.*/
:SOUR:CURR:VON? /*Queries the starting voltage (Von) set in CC mode.*/
[:SOURce]:CURRent:VLIMt
Syntax [:SOURce]:CURRent:VLIMt {<value>|MINimum|MAXimum|DEFault}
[:SOURce]:CURRent:VLIMt? [MINimum|MAXimum|DEFault]
Description Sets the voltage limit in CC mode.
Queries the voltage limit set in CC mode.
Remarks The voltage limit refers to the upper limit of the voltage working in CC mode. The unit
is V.
In the setting command, the parameter "MINimum" indicates setting the minimum
voltage; the parameter "MAXimum" indicates setting the maximum voltage; and the
parameter "DEFault" indicates setting the default voltage limit.
In the query command, the parameter "MINimum" indicates querying the minimum
voltage; the parameter "MAXimum" indicates querying the maximum voltage; and the
parameter "DEFault" indicates querying the default voltage limit value.
Return
The query returns a real number.
Example :SOUR:CURR:VLIM 5 /*Sets the voltage limit in CC mode to be 5 V.*/
:SOUR:CURR:VLIM? /*Queries the voltage limit set in CC mode.*/
[:SOURce]:CURRent:ILIMt
Syntax [:SOURce]:CURRent:ILIMt {<value>|MINimum|MAXimum|DEFault}
[:SOURce]:CURRent:ILIMt? [MINimum|MAXimum|DEFault]
Description Sets the current limit in CC mode.
Queries the current limit set in CC mode.