Chapter 2 Command System RIGOL
DL3000 Programming Guide 2-47
mode. The default unit for the value is W.
In the setting command, the parameter "MINimum" indicates setting the minimum
value of the maximum power; the parameter "MAXimum" indicates setting the
maximum value of the maximum power; and the parameter "DEFault" indicates
setting the default value of the maximum power.
In the query command, the parameter "MINimum" indicates querying the minimum
value of the maximum power; the parameter "MAXimum" indicates querying the
maximum value of the maximum power; and the parameter "DEFault" indicates
querying the default value of the maximum power.
Return
The query returns a real number.
Example
:SOUR:OPP:PMAX 100 /*Sets the maximum power in OPP mode to be 100 W.*/
:SOUR:OPP:PMAX? /*Queries the maximum power in OPP mode.*/
[:SOURce]:OPP:PMIN
Syntax [:SOURce]:OPP:PMIN {<value>|MINimum|MAXimum|DEFault}
[:SOURce]:OPP:PMIN? [MINimum|MAXimum|DEFault]
Description Sets the minimum power in OPP mode.
Queries the minimum power in OPP mode.
Remarks The minimum power is called the minimum value of the protection power in the OPP
mode. The default unit for the value is W.
In the setting command, the parameter "MINimum" indicates setting the minimum
value of the minimum power; the parameter "MAXimum" indicates setting the
maximum value of the minimum power; and the parameter "DEFault" indicates
setting the default value of the minimum power.
In the query command, the parameter "MINimum" indicates querying the minimum
value of the minimum power; the parameter "MAXimum" indicates querying the
maximum value of the minimum power; and the parameter "DEFault" indicates
querying the default value of the minimum power.
Return
The query returns a real number.
Example
:SOUR:OPP:PMIN 90 /*Sets the minimum power in OPP mode to be 90 W.*/
:SOUR:OPP:PMIN? /*Queries the minimum power in OPP mode.*/
[:SOURce]:OPP:VOPP
Syntax [:SOURce]:OPP:VOPP {<value>|MINimum|MAXimum|DEFault}
[:SOURce]:OPP:VOPP? [MINimum|MAXimum|DEFault]
Description Sets the protection voltage in OPP mode.
Queries the protection voltage in OPP mode.
Remarks The default unit for the protection voltage is V.
In the setting command, the parameter "MINimum" indicates setting the minimum
protection voltage; the parameter "MAXimum" indicates setting the maximum
protection voltage; and the parameter "DEFault" indicates setting the default