Chapter 2 Command System RIGOL
DL3000 Programming Guide 2-37
In the setting command, the parameter "MINimum" indicates setting a low range;
the parameter "MAXimum" indicates setting a high range; and the parameter
"DEFault" indicates setting the default range.
In the query command, the parameter "MINimum" indicates querying a low range;
the parameter "MAXimum" indicates querying a high range; and the parameter
"DEFault" indicates querying the default range.
Return
Format
The query returns a real number.
Example
:SOUR:BATT:RANG 60 /*Sets the current range in Battery mode to be 60 A (a high
range).*/
:SOUR:BATT:RANG? /*Queries the current range set in Battery mode.*/
[:SOURce]:BATTary[:LEVel][:IMMediate]
Syntax [:SOURce]:BATTary[:LEVel][:IMMediate] {<value>|MINimum|MAXimum|DEFault}
[:SOURce]:BATTary[:LEVel][:IMMediate]?[MINimum|MAXimum|DEFault]
Description Sets the load regulation current in Battery mode.
Queries the load regulation current set in Battery mode.
Remarks The load regulation current is called the discharge current in Battery mode. Its
default unit is A.
In the setting command, the parameter "MINimum" indicates setting the minimum
load regulation current; the parameter "MAXimum" indicates setting the maximum
load regulation current; and the parameter "DEFault" indicates setting the default
load regulation current.
In the query command, the parameter "MINimum" indicates querying the minimum
load regulation current; the parameter "MAXimum" indicates querying the maximum
load regulation current; and the parameter "DEFault" indicates querying the default
load regulation current.
Return
Format
The query returns a real number.
Example
:SOUR:BATT:LEV:IMM 3 /*Sets the load regulation current in Battery mode to be 3
A.*/
:SOUR:BATT:LEV:IMM? /*Queries the load regulation current in Battery mode.*/
[:SOURce]:BATTary:VSTop
Syntax [:SOURce]: BATTary:VSTop {<value>|MINimum|MAXimum|DEFault}
[:SOURce]: BATTary:VSTop? [MINimum|MAXimum|DEFault]
Description Sets the voltage at which the battery stops discharging in Battery mode.
Queries the voltage at which the battery stops discharging in Battery mode.
Remarks The default unit for the voltage is V.
In the setting command, the parameter "MINimum" indicates setting the minimum
voltage; the parameter "MAXimum" indicates setting the maximum voltage; and the
parameter "DEFault" indicates setting the default voltage.
In the query command, the parameter "MINimum" indicates querying the minimum
voltage; the parameter "MAXimum" indicates querying the maximum voltage; and

Table of Contents

Other manuals for Rigol DL3000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DL3000 and is the answer not in the manual?

Rigol DL3000 Specifications

General IconGeneral
BrandRigol
ModelDL3000
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals