RIGOL Chapter 2 Command System
2-44 DL3000 Programming Guide
In the query command, the parameter "MINimum" indicates querying the minimum
protection voltage; the parameter "MAXimum" indicates querying the maximum
protection voltage; and the parameter "DEFault" indicates querying the default
Return
The query returns a real number.
Example
:SOUR:OCP:VOCP 1 /*Sets the protection voltage in OCP mode to be 1 V.*/
:SOUR:OCP:VOCP? /*Queries the protection voltage in OCP mode.*/
[:SOURce]:OCP:TOCP
Syntax [:SOURce]:OCP:TOCP {<value>|MINimum|MAXimum|DEFault}
[:SOURce]:OCP:TOCP? [MINimum|MAXimum|DEFault]
Description Sets the maximum overcurrent protection time in OCP mode.
Queries the maximum overcurrent protection time in OCP mode.
Remarks When the maximum overcurrent protection time is exceeded and no action is
performed, the load will automatically stop sinking the current. The default unit is
μs.
In the setting command, the parameter "MINimum" indicates setting the minimum
protection time; the parameter "MAXimum" indicates setting the maximum
protection time; and the parameter "DEFault" indicates setting the default protection
time.
In the query command, the parameter "MINimum" indicates querying the minimum
protection time; the parameter "MAXimum" indicates querying the maximum
protection time; and the parameter "DEFault" indicates querying the default
Return
The query returns a real number.
Example
:SOUR:OCP:TOCP 500 /*Sets the maximum overcurrent protection time in OCP
mode to be 500
μs.*/
:SOUR:OCP:TOCP? /*Queries the maximum overcurrent protection time in
[:SOURce]:OPP:VON
Syntax [:SOURce]:OPP:VON {<value>|MINimum|MAXimum|DEFault}
[:SOURce]:OPP:VON? [MINimum|MAXimum|DEFault]
Description Sets the starting voltage in OPP mode.
Queries the starting voltage in OPP mode.
Remarks The default unit for the starting voltage is V, and it ranges from 0 V to 150 V.
In the setting command, the parameter "MINimum" indicates setting the minimum
starting voltage; the parameter "MAXimum" indicates setting the maximum starting
voltage; and the parameter "DEFault" indicates setting the default starting voltage.
In the query command, the parameter "MINimum" indicates querying the minimum
starting voltage; the parameter "MAXimum" indicates querying the maximum
starting voltage; and the parameter "DEFault" indicates querying the default starting